Ultrafiltration Tubes: Where 90% of Failed Runs Go Wrong
15 mL · Cat. No. JKR-CM-15 · Protein concentration / buffer exchange / desalting
A single ultrafiltration tube can take 10 mL of precious sample down to 200 μL in half an hour — or ruin a week of culture through one overlooked detail. We distilled the parts of our user guide that are most often skipped and most costly to get wrong. Save this and check it against your next run.
1. Choose Right: MWCO Decides the Outcome
Too low a molecular weight cut-off (MWCO) and concentration crawls; too high and your protein passes straight through. The rule of thumb: target molecular weight ≥ MWCO × 3. Use 30 or 50 kDa for IgG (150 kDa), 100 kDa for IgM (900 kDa), and 30 kDa for typical enzymes (50–100 kDa). Where recovery matters, test two adjacent cut-offs in a pilot run — globular proteins and linear peptides do not retain alike.
2. Always First: Pre-Rinse and Leak Check
PES membranes ship with trace glycerol and other pore-keeping agents that will contaminate your sample if not rinsed away. Add 12 mL (fixed-angle) or 15 mL (swing-bucket) of buffer or ultrapure water, cap tightly, invert for 5–10 s to check for leakage, then spin at 2000–2500 × g for 5–10 min and discard. This confirms the seal and wets the membrane for optimal flux. Once wet, the membrane must stay wet — leave buffer on it and store at 4 °C if not used immediately.
3. Five Red Lines You Cannot Cross
|
Spin-force limit |
≤ 4000 × g recommended, never above 5000 × g. Overspeeding risks pass-through; rupture loses the sample and contaminates the whole rotor. |
|
Balancing |
Keep the weight difference ≤ 0.1 g using a balance. Imbalance is the leading cause of tube failure. |
|
Sample prep |
Filter through 0.22 μm or centrifuge out particles, precipitates, and aggregated protein first — otherwise the membrane fouls irreversibly. |
|
Never dry out |
A dry membrane changes pore structure and recovery collapses — practically scrap. |
|
No tip contact |
Puncturing the membrane with a pipette tip ends the run instantly. Load and aspirate along the wall. |
4. The Step That Lifts Recovery from 80% to 95%
The last 50–100 μL in the upper chamber is the hardest to retrieve. After the final spin, add 100–200 μL of buffer, pipette gently a few times, then withdraw — recovery rises from 80% to above 95%. Two more habits pay off: keep the starting protein concentration at 1 mg/mL or higher to limit membrane adsorption, and hold the filtrate until the run is verified before discarding it.
5. Desalting and Exchange: Three Rounds Minimum
Concentrate to one-third to one-half of the original volume, discard the filtrate, refill with an equal volume of target buffer, mix, and concentrate again. Repeat at least three times for over 99% exchange. Keep the pH difference between starting and final buffer within 1 unit, or protein may precipitate mid-run. For high-salt samples (> 500 mM), dilute 2–3 fold first. Run at 4–10 °C: it protects labile protein and lowers viscosity.
6. Can It Be Reused?
Plainly: this is a non-sterile single-use device, and single use is our official recommendation. Publication work, clinical studies, and GMP/GLP settings require one-time use with retained samples for traceability. Only in exploratory, non-publication work with cross-contamination risk assessed may a tube be reused within one protein and one buffer system, up to 10 cycles — discard once flux falls to 60% of initial. Cleaning: soak in 0.1 M NaOH for 20 min at room temperature → rinse three times with Milli-Q water → soak in 20% ethanol for 4 h or overnight → store wet at 4 °C. Never autoclave, UV-irradiate for extended periods, or soak in alcohol, and avoid oxidants such as sodium hypochlorite; all of these damage the PES membrane.
7. Keep on File
Unopened devices keep 3 years at room temperature, dry, out of direct sun, under 70% RH, stacked no more than 5 high. After opening, reseal the bag and use within 3 months. Cleaned and reused tubes keep no longer than 7 days (4 °C, sealed, wetted with 20% ethanol or buffer). A built-in dead-stop at about 200 μL prevents over-concentration; most samples finish filtering in 20 min and reach minimum volume in 30–40 min.
| Slow flux, leakage, or rupture — or need custom MWCO, sterile pre-packed, or bulk formats? GeneCreate technical support: +86-27-87960366 | marketing@genecreate.com | www.genecreate.cn For research use by professionals only. Not for clinical diagnosis or treatment. Wear a lab coat and disposable gloves.
|
Adapted from "Ultrafiltration Centrifugal Tube — Usage Notes V1.0". General guidance only; optimize parameters for your own samples.